Troubleshooting SPrint
Each day, before you turn the processor on, check that the cable is properly connected to the processor and not broken or cracked; the sensitivity or volume control level is as recommended by your clinician; and the transmitting coil is in place.
Each night, after you turn the processor off, remove the batteries, and store the speech processor overnight with the drying aid. Moisture or humidity may cause cut-outs or failures.
Follow these steps
If you cannot hear any sound from your processor, follow these steps:
1. Replace the batteries with fresh ones if the LCD panel is not displayed, or if the Low Battery symbol is showing.
2. check the batteries are correctly inserted. Occasionally clean the battery contacts, with the cleaning brush.
rechargeable batteries may need to be replaced if they have expired or the battery charger is not recharging them.
3. With the SPrint™ turned on, place the transmitting coil with the cables still attached to the SPrint,
over the back cover of the processor. The LCD panel should show the RF test coil symbol.
4. Alternatively, use the Signal Check to check the transmitting coil. If the Signal Check lights up, the coil is functioning. If it does not light up, replace the transmitting coil, transmitting cable or headset cable.
When the Manual and Service spanner symbols are shown at the same time on the LCD panel:
1. Reset the processor, by turning it OFF and then ON again.
2. Select another program by using the program switch button. In some cases this will be automatically selected. Contact your clinic.
3. Ask a hearing person to use monitor earphones to check the microphone.
4. Attach and use the lapel microphone. An improvement in sound quality with this microphone indicates the SPrint microphone may be faulty or may need drying out with the drying aid. Contact your clinic if drying out does not solve the problem.
5. Try moving away from any electronic device that may be causing interference.
After each step, check if there is any sound. If there is still no sound, contact your clinician, cochlear implant center or your nearest Cochlear office.
- Support
- Baha
- Cochlear Implants
- Accessories, ALDs
- Back to School with CI
- Candidacy Criteria
- Cochlear Implant Batteries
- FM Options
- Fire Safety Tips
- Frequently Asked Questions
- Getting to Know Your Processor
- Graeme Clark Scholarship
- Hear Always - product repair
- MRI & Other Questions
- Music - enhance music enjoyment
- Nucleus 5 Instructional Videos
- Nucleus CR110 Remote Assistant Interactive Guide
- Organizations and Support
- Service Agreements
- SkinIt Processor covers
- Sports and Summer Activities
- Surgery Preparation
- Telephone and Cell Phone Tips
- Travel Tips
- Troubleshooting for Nucleus Sound Processors
- Upgrade Program
- User Manuals and Guides
- Zoom Technology
- Customer Service